HA-8160及DS5测定糖化血红蛋白的比对分析及偏倚评估

董磊 刘娟 马红雨 全首祯 朱美财

现代检验医学杂志2012,Vol.27Issue(6):100-101,104,3.
现代检验医学杂志2012,Vol.27Issue(6):100-101,104,3.DOI:10.3969/j.issn.1671-7414.2012.06.031

HA-8160及DS5测定糖化血红蛋白的比对分析及偏倚评估

Comparison Analysis and Bias Evalution of HA-8160 and DS5 Automatic Glycosylated Hemoglobin Analyzer

董磊 1刘娟 2马红雨 1全首祯 1朱美财1

作者信息

  • 1. 空军总医院临床检验中心,北京,100142
  • 2. 空军总医院输血科,北京,100142
  • 折叠

摘要

Abstract

Objective To compare the results of HbAlc detected by HA-8160 and DS5 production of ARKRAY of Japan and DREW of England, respectively. Methods The precision and carryover rates were evaluated. Blood samples of 40 patients were examined by HA-8160 and DS5. Results All the precision and carryover rates were within an acceptable range. There was no significant difference in the HbAlc results between HA-8160 and DS5,and the bias in two machines was within an acceptable range. Conclusion The precision and carryover rates were according with the damands of the experiments in clinical diagnosis. The HbAlc results detected by HA-8160 and DS5 were comparable which both could provide reliable datas.

关键词

糖化血红蛋白分析仪/比对/偏倚

Key words

glycosyiated hemoglobin analyzer/comparison/bias