Abstract
Objective To compare the characteristics of radiophysics of the muti-fields full body irradiation by 4 Mev and 6 Mev mixed-energy electron beam with the 4 Mev or 6 Mev single-energy electron beam. Methods 3D-water tank, anthropomorphic phantom and dosimeter were used to measure the physical and clinical parameters of 4 Mev or 6 Mev single-energy electron beam or 4 Mev and 6 Mev mixed-energy electron beams generated by Elektra Precise Accelerator, respectively. In addition, the actual measurement results were summarized, analyzed and compared with each other so as to work out the optimal energy allocation of physical and clinical parameters for treating mycosis fungoides. Results When a 5 mm plexiglass scattering screen was applied, the actually measured physical parameters of the 4 Mev and 6 Mev mixed-energy electron beam could meet the clinical requirements. Conclusion The mixed-energy irradiation can not only meet the clinical treatment requirements, but also shorten the treatment time compared to the single-energy electron beam radiation treatment.关键词
全身电子线照射(TSEI)/混合能量/剂量/蕈样霉菌病Key words
total skin electron irradiation (TSEI)/ complex energy/ dose/ mycosis fungoides分类
医药卫生引用本文复制引用
钱建升,张晓智,田美丽..混合能量电子束全身照射技术的临床应用及特点[J].西安交通大学学报(医学版),2013,34(1):122-124,3.