摘要

Abstract

Objective To introduce the quality control (QC) testing of 18 defibrillation ECG monitors in a Level 3 hospital in Wuxi. Methods Based on measurement standards of Zhejiang province, we tested the electrical safety, energy output and electrocardiogram monitor using FLUKE electric analyzer and defibrillator analyzer. Results Qualified rates are 100% except delay time under synchronous mode. Conclusion By QC testing, we hope to reduce the use nsk of defibritlators, improve medical quality, and further focus on the detection of defibrillator battery and risk control of delay time under synchronous mode.

关键词

除颤心电监控仪/除颤仪/质量控制/电气安全分析仪

Key words

defibrillation ECG monitor/ defibrillator/ quality control/ electrical safety analyzer