| 注册
首页|期刊导航|中国医学装备|256层CT前瞻性心电门控冠状动脉成像研究

256层CT前瞻性心电门控冠状动脉成像研究

王继芳 姚以刚 徐学权 李春林 任科研 马天赐

中国医学装备2013,Vol.10Issue(2):74-76,3.
中国医学装备2013,Vol.10Issue(2):74-76,3.

256层CT前瞻性心电门控冠状动脉成像研究

Sudy of prospective ECG-triggered for 256-slice CT coronary artery angiography

王继芳 1姚以刚 1徐学权 1李春林 1任科研 1马天赐1

作者信息

  • 1. 辽宁省铁法煤业集团总医院放射科CT室,辽宁调兵山,112700
  • 折叠

摘要

Abstract

Objective: To compare the image quality and patient radiation dose between prospective ECG-triggered and retrospective ECG-gated CT coronary angiography (CTCA) by using 256-slice CT. Methods: A total of 200 patients were enrolled in this study. All patients with an average heart rate (HR) <70 bpm, 100 patients were scanned by using prospective ECG-triggered and 100 patients by retrospective ECG-gated with 256-slice CT, The image quality and the radiation dose were compared in two groups. Results: of 800 coronary artery, excellent or good image quality was achieved in 96.2°/o(385/400) in group A, 97.3°/o(389/400) in group B, with no significant difference for A vs B. The average radiation dose in group A [(2.9±0.5)mSv] reduced 73% comparing with that in group B[(10.2±2.4)mSv]. Conclusion: Using prospective ECG-triggered to perform CCTA with 256-slice CT could keep the image quality and reduce radiation dose significantly.

关键词

体层摄影术,X线计算机/冠状血管/辐射剂量

Key words

Tomography/ X-ray computde/ Coronary vessel/ Radiation dosage