计量学报2013,Vol.34Issue(1):76-78,3.DOI:10.3969/j.issn.1000-1158.2013.01.17
开环式测量氡析出率
Measurement of Radon Exhalation Rate in Open Loop
摘要
Abstract
The measurement system for radon exhalation rate in open loop is validated. It shows that the radon exhalation rate from medium surface obtained is very good to match the reference value. The time expression when the radon concentration in the chamber tends to a constant value is obtained by theoretical analysis. A method is proposed that a chamber with bigger bottom acreage is used and a adjust valve is added to lower the flux of pump, such increasing the radon concentration in the chamber and reducing the statistical fluctuation in the measurement with RAD7.关键词
计量学/氡析出率/开环Key words
Metrology/Radon exhalation rate/Open loop分类
通用工业技术引用本文复制引用
谭延亮,肖德涛,赵桂芝,周青芝,肖田,周毅..开环式测量氡析出率[J].计量学报,2013,34(1):76-78,3.基金项目