GE Definium 6000 DR图像白边伪影消除

陈明 杭晓鹏 高文峰

中国医学装备2013,Vol.10Issue(3):93-94,2.
中国医学装备2013,Vol.10Issue(3):93-94,2.

GE Definium 6000 DR图像白边伪影消除

Removal of white edge fault in GE Definium 6000 DR images

陈明 1杭晓鹏 1高文峰1

作者信息

  • 折叠

摘要

Abstract

Figure out the GE Definium 6000 DR image of the white edge artifacts. Detector calibration and Shu Guangqi light position adjustment. Elimination of the image of the white edge artifacts. The current domestic medical equipment repair, due to technical limitations, basically is the repair, and the factory repair using a modular repair, replacement of the whole component or circuit board, the high cost of.But in the practical work, and not every equipment failures are difficult, if our engineers can in the event of a fault, according to the information and repair experience, self checking, possibly through simple method of problem solution.

关键词

数字化X射线摄影/平板探测器/伪影/校准

Key words

Digital X ray photography/ Flat panel detector/ Artifact/ Calibration