中国医学装备2013,Vol.10Issue(3):93-94,2.
GE Definium 6000 DR图像白边伪影消除
Removal of white edge fault in GE Definium 6000 DR images
摘要
Abstract
Figure out the GE Definium 6000 DR image of the white edge artifacts. Detector calibration and Shu Guangqi light position adjustment. Elimination of the image of the white edge artifacts. The current domestic medical equipment repair, due to technical limitations, basically is the repair, and the factory repair using a modular repair, replacement of the whole component or circuit board, the high cost of.But in the practical work, and not every equipment failures are difficult, if our engineers can in the event of a fault, according to the information and repair experience, self checking, possibly through simple method of problem solution.关键词
数字化X射线摄影/平板探测器/伪影/校准Key words
Digital X ray photography/ Flat panel detector/ Artifact/ Calibration分类
医药卫生引用本文复制引用
陈明,杭晓鹏,高文峰..GE Definium 6000 DR图像白边伪影消除[J].中国医学装备,2013,10(3):93-94,2.