血液透析器不良事件故障树模型的建立
Establishment of Fault Tree Model of Dialyzer Adverse Events
摘要
Abstract
Objective To establish the fault tree model of dialyzer adverse events in order to provide a theoretical guide and reference for traceability analysis of adverse events in high -value consumables such as dialyzers. Methods Fault Tree Analysis (FTA) method was used in risk analysis of dialyzers through collecting adverse events occurring in two hospitals' blood purification departments in Shanghai from June 2012 to August 2012, with theoretical knowledge and practical experience combined. Results The various types of dialyzers adverse events could be described comprehensively by the fault tree model. Conclusion The establishment of dialyzer adverse events analysis model can provide a reference for the traceability of quality control, also can improve management efficiency and promote the commencement of the risk management activities.关键词
血液透析器/不良事件/故障树分析法Key words
dialyzer/ adverse events/ fault tree analysis(FTA) method分类
医药卫生引用本文复制引用
谢培杰,张斌,张飞鸿,盛晓华,赵丽萍,计雁飞..血液透析器不良事件故障树模型的建立[J].医疗卫生装备,2013,34(3):7-10,4.基金项目