2种电磁屏蔽材料在3.0T磁共振成像系统的应用探讨

吉善创 黄邻彬 叶伟

医疗卫生装备2013,Vol.34Issue(3):101-102,2.
医疗卫生装备2013,Vol.34Issue(3):101-102,2.DOI:10.7687/J.ISSN.1003-8868.2013.03.101

2种电磁屏蔽材料在3.0T磁共振成像系统的应用探讨

Application Study of Two Electromagnetism Shielding Materials in 3.0T Magnetic Resonance Imaging System

吉善创 1黄邻彬 1叶伟1

作者信息

  • 1. 广西医科大学第一附属医院,南宁530021
  • 折叠

摘要

Abstract

Objective To approach the application of red copper coin and cold-rolled galvanized steel as the electromagnetic shielding materials in 3.0T magnetic resonance imaging system. Methods Phillips Achiva 3.0T magnetic resonance imaging system used red copper coin, a kind of nonmagnetic material, as the shielding material, and Siemens Verio 3.0T magnetic resonance imaging system employed cold-rolled galvanized steel, a kind of soft magnetic material, as the shielding material. After the main magnet was hoisted into the scanning room, the nick of shielding was closed. The shielding effectiveness was detected before excitation. Results The measurement was 100 dB to 106 dB for the red copper coin with a mean value of (102±2)dB and 101 dB to 106 dB for the cold-rolled galvanized steel with a mean value of (103±l)dB. There was no significant difference between the performances of the two materials with P>0.05, while magnetic lines might penetrate the red copper coin and the closed hexahedron structure behaved not so good as that of cold-rolled galvanized steel. Conclusion There is no significant between the shielding effectiveness of the two materials, while red copper shielding room may have some magnetic lines fleeing to the external wall.

关键词

电磁屏蔽/磁共振/超导磁共振

Key words

electromagnetic shielding/ magnetic resonance/ superconduction magnetic resonance

分类

医药卫生

引用本文复制引用

吉善创,黄邻彬,叶伟..2种电磁屏蔽材料在3.0T磁共振成像系统的应用探讨[J].医疗卫生装备,2013,34(3):101-102,2.

医疗卫生装备

OACSTPCD

1003-8868

访问量0
|
下载量0
段落导航相关论文