辐射防护2013,Vol.33Issue(3):185-188,4.
高剂量率后装放疗辐射场的最优化研究
Study on Optimization of Rradiation Field for High Dose Rate Afterloading Radiotherapy
邓君 1曹磊 1王晓刊 1王成国 2林志凯 1王建超 1练德幸1
作者信息
- 1. 中国疾病预防控制中心辐射防护与核安全医学所,辐射防护与核应急中国疾病预防控制中心重点实验室,北京100088
- 2. 内蒙古疾病预防控制中心,呼和浩特010031
- 折叠
摘要
Abstract
Based on a high dose rate afterloading radiotherapy,the γ radiation level at inner entrance of the maze and the first transition in inner entrance of the maze as well as the outer entrance of the maze were calculated,and the influence of source' s location to the distribution of radiation level in the maze was analyzed through a monte-carlo code,whose reliability was tested by comparing with the results measured on-site.Results show that the radiation level in the maze strongly depends on the location of radioactive source in the treat room,and reach to a relatively lower level when the source locates at the east of treat room and close to the maze wall which can be recommended as an optimized location for source when executing the therapy.关键词
高剂量率(HDR)后装放疗/MC程序/最优化Key words
high dose rate afterloading radiotherapy/ monte-carlo code/ optimization分类
医药卫生引用本文复制引用
邓君,曹磊,王晓刊,王成国,林志凯,王建超,练德幸..高剂量率后装放疗辐射场的最优化研究[J].辐射防护,2013,33(3):185-188,4.