不同剂量放射源对Infinia Hawkeye均匀性校正的影响

刘金华 黄柳华 冼嘉朗 廖玉娥 冯会娟 胡瑞 孙云钢 欧阳伟

医疗卫生装备2013,Vol.34Issue(7):113-114,2.
医疗卫生装备2013,Vol.34Issue(7):113-114,2.DOI:10.7687/J.ISSN.1003-8868.2013.07.113

不同剂量放射源对Infinia Hawkeye均匀性校正的影响

Influences of Radioactive Sources with Different Doses on Infinia Hawkeye Uniformity Calibration

刘金华 1黄柳华 1冼嘉朗 1廖玉娥 1冯会娟 1胡瑞 1孙云钢 1欧阳伟1

作者信息

  • 1. 南方医科大学珠江医院核医学科,广州510282
  • 折叠

摘要

Abstract

Objective To determine the influences of the radioactive sources with different doses on Infinia Hawkeye uniformity correction.Methods The radioactive sources with different doses were prepared,and underwent uniformity calibration test respectively.Results The uniformity calibrations by different sources were not the same,and also the time expended.Conclusion The dose of 20 kcts/s or 30 kcts/s is ideal for the radioactive source.

关键词

Infinia Hawkeye/放射源/均匀性/不同剂量

Key words

Infinia Hawkeye/ radioactive source/ uniformity/ different doses