Abstract
This paper studied the uncertainty of the RAD7 emanometer during measuring Ra in air.By simulating different concentration of Ra in standard Ra chamber,the uncertainty of the emanometer was studied in standard conditions.The uncertainties were 14.38% and 14.52% in different Ra concentration when k=2.This RAD7 can reach the demands of the determination uncertainty of Ra in air of the GB50325-2001.关键词
空气/氡/测量/不确定度Key words