摘要
Abstract
Objective To explore the results consistency of ion(K,Na,Cl,Ca,CO2) detection between the self-built detection system and supporting system which is as the target system after value transfer.Methods According to EP9-A2 document of Committee of Clinical Laboratory Standards(NCCLS),the supporting detection system was composed of BECKMAN CX5 automatic biochemical analyzer and its original reagents and calibrator as the target system,and the self-built system was composed of Toshiba TBA-120 automatic biochemical analyzer and RANDOX company reagent as the assay detection system.Supporting system gave the fresh serum assignment which was as the calibrator to calibrate the self-built system.Mter value transfer,such items as K,Na,Cl,Ca,CO2 were detected by using the fresh serum from patients and the deviation of the two systems was evaluated.Results Deviations of the results of five detected items were within the allowable range and acceptable.Conclusion It's an effective way to solve the results consistency between the self-built detection system and supporting detection system by value transfer using fresh serum from patients.关键词
自建离子检测系统/量值传递/比对/配套检测系统Key words
self-built ion detection system/ value transfer/ comparison/ supporting detection system分类
医药卫生