首页|期刊导航|中国医学装备|Mindray BC-5300自动血细胞分析仪性能的初步评价

Mindray BC-5300自动血细胞分析仪性能的初步评价

徐菡 乐家新 兰亚婷 任爽

中国医学装备2013,Vol.10Issue(8):6-9,4.
中国医学装备2013,Vol.10Issue(8):6-9,4.DOI:10.3969/J.ISSN.1672-8270.2013.08.002

Mindray BC-5300自动血细胞分析仪性能的初步评价

Performance evaluation of mindray BC-5300 hematology analyzer

徐菡 1乐家新 1兰亚婷 1任爽1

作者信息

  • 1. 解放军总医院临床检验科 北京 100853
  • 折叠

摘要

Abstract

Objective:To evaluate the performance of Mindray BC-5300 hematology analyzer on completing blood cell counting(CBC) and leukocyte differential count(DC).Methods:The indicators of CBC performance were evaluated by following guidelines and compared with manual operation.Results:The precision,carry-over rate,and total repeatability meet the standards,and the linearity was also reliable both in physiogenic and pathological conditions.Comparing with the results from XE-2100,the correlation was also satisfied(r2>0.994).It has better repeatability in neutrophilic granulocytes,lymphocytes,eosinophile granulocyte,and basophile granulocyte.The WBC samples with morphological performance were compared between instrumental and manual DC operation.The repeatability in neutrophilic granulocytes,lymphocytes and eosinophile granulocyte(r=0.9663,r=0.9595,r=0.9178),the r value of monocyte was 0.9104 and basophil was 0.5065.The detection of juvenile cell and abnormal lymphocyte suggested manual review.Conclusion:The results from BC-5300 automatic hematology analyzer were accurate and reliable,the main indicators could all meet the clinical needs.The suggestive diagnosis by leukocyte scatterplot and abnormal leukocytes classification should be double checked by manual smear microscopic examination.

关键词

血细胞分析仪/精密度/重复性/携带污染率/线性

Key words

Hematology analyzer/ Precision/ Reproducibility/ Carryover/ Linearity

引用本文复制引用

徐菡,乐家新,兰亚婷,任爽..Mindray BC-5300自动血细胞分析仪性能的初步评价[J].中国医学装备,2013,10(8):6-9,4.

基金项目

国家科技支撑计划课题(2013BAI17B05)"基层医院血细胞分析应用评价研究" (2013BAI17B05)

中国医学装备

OACSTPCD

1672-8270

访问量0
|
下载量0
段落导航相关论文