国际眼科杂志Issue(12):2551-2553,3.DOI:10.3980/j.issn.1672-5123.2013.12.59
前节OCT和pentacam及A超在测量角膜厚度上的比较
Comparison of central corneal thickness measured by anterior segment optical coherence tomography, pentacam and ultrasonic pachymetry
赵炜 1吴婷 2董泽红 1冯洁 1任玉凤 1王雨生1
作者信息
- 1. 710032 中国陕西省西安市,第四军医大学西京医院眼科全军眼科研究所
- 2. 710025 中国陕西省西安市,第二炮兵工程大学门诊部
- 折叠
摘要
Abstract
AIM: To investigate and analyze the differences in central corneal thickness ( CCT ) measurement by anterior segment optical coherence tomography ( AS-OCT ) , pentacam rotating Scheimpflug camera and ultrasonic pachymetry ( USP) . <br> METHODS: The retrospective study measured CCT of 43 patients ( 86 eyes ) performed excimer laser myopia corrective surgery by the same examiner using RTVue AS-OCT with an anterior segment adaptor, pentacam rotating Scheimpflug camera, and PACSCAN300P USP. Data of three groups were compared and statistically analyzed by pair t test. <br> RESULTS: The mean CCT ( ±SD ) measured by AS-OCT, pentacam and USP were 535 ±52, 550 ±52, 550 ± 51μm, respectively.Paired-sample t test showed that the results of AS -OCT, pentacam and USP were statistically different. There were no statistically significant differences in the mean CCT between the pentacam and USP group. Significant correlations of measurement result were found between pentacam and USP (r=0.998, P<0.01), FD-OCT and USP (r=0.993, P<0.01), and AS-OCT and pentacam (r=0.991, P<0.01). <br> CONCLUSION:RTVue AS -OCT significantly under estimates the CCT, while the measurement results by pentacam are similar to those by USP which is regarded as the golden standard.关键词
前节OCT/三维眼前节分析仪/A超/角膜厚度/测量Key words
anterior segment optical coherence tomography/pentacam/ultrasonic pachymetry/central corneal thickness/measure引用本文复制引用
赵炜,吴婷,董泽红,冯洁,任玉凤,王雨生..前节OCT和pentacam及A超在测量角膜厚度上的比较[J].国际眼科杂志,2013,(12):2551-2553,3.