摘要

Abstract

Objective: To solve some common problems in the verification process of digital electrocardiograph through non-maintenance solutions. Methods Verifying digital electrocardiograph with EGC-1C tester according to verification regulation JJG1041-2008, and analyzing problems occurred in the veriifcation process to ifnd corresponding solutions. Results Setting problems are settled by changing the settings of digital electrocardiograph. Design problems can be settled by consulting relevant experts or manufacturers. Conclusion Digital electrocardiograph must be veriifed according to the veriifcation regulation. The instruments which have some problems discovered by metrological verification don’t neet to be repaired immediately. We should analyze the problems firstly to find corresponding solutions.

关键词

数字心电图机/计量检定/内定标电压/加权系数/幅度测量/幅频特性/交流干扰

Key words

digital electrocardiograph/metrological verification/calibration voltage/weighting coefifcient/amplitude measurement/amplitude frequency characteristics/alternating current interference