FDP试剂在STA-R全自动血凝分析仪上使用的性能评估

The Evaluation of the Analytical Performance of the Reagent for the FDP (Nanopia P-FDP produced by SEKISUI MEDICAL CO. LTD) on STA-R

郑苗 1倪琛 1薛晓光1

作者信息

  • 1. 福建省妇幼保健院,福建 福州 350001
  • 折叠

摘要

Abstract

Objective To evaluate the analytical performance of the reagent for the FDP (Nanopia P-FDP produced by SEKISUI MEDICAL CO., LTD) on STA-R. Methods The precision, the linearity and the frontal zone phenomenon were evaluated. Results The imprecision of intra-assay was 2.8%~7.3%, two dif erent control items and pooled plasma checked. The linearity range of the reagent was 0.00~136ug/mL. There is no failed results in the test of frontal zone phenomenon, when carry on the over-range material (936.83ug/mL). When the sample is high at FDP (>467.80ug/mL), STA-R makes a Alarm (V>Vmax). Conclusions The comparison test shows the method for detecting FDP was excellent, Nanopia P-FDP on STA-R is suitable for FDP measurement. The method can be used in clinical diagnosis.

关键词

纤维蛋白原/纤维蛋白降解产物/试剂盒/评价研究

Key words

FDP/Reagent kit/Evaluation studies/Nanopia P-FDP