原子能科学技术Issue(3):523-527,5.DOI:10.7538/yzk.2014.48.03.0523
60 Co 单源照射装置的蒙特卡罗方法优化设计
Optimal Design of 60 Co Single Source Radiation Facility With Monte Carlo Method
高飞 1肖雪夫 1侯金兵 1王红玉 1宋明哲 1倪宁 1张力1
作者信息
- 1. 中国原子能科学研究院 放射性计量测试部,北京 102413
- 折叠
摘要
Abstract
Reference radiation is necessary for calibration of radiation dose monitoring meters ,and 60 Co single source radiation facility is an important equipment to produce reference radiation field .Monte Carlo method was used to calculate the uniformity and scattered radiation of the reference radiation field which was produced by 60Co single source radiation facility ,and the design of scattered chamber ,collimation and shutter was accomplished .PTW ionization chamber was used to measure the radiation field ,and the results show that the scattered radiation and uniformity of the radiation field are in good agreement with the simulation results .The radiation performances meet the design requirements .关键词
单源照射装置/蒙特卡罗/参考辐射场/散射辐射Key words
single source radiation facility/Monte Carlo/reference radiation field/scat-tered radiation分类
能源科技引用本文复制引用
高飞,肖雪夫,侯金兵,王红玉,宋明哲,倪宁,张力..60 Co 单源照射装置的蒙特卡罗方法优化设计[J].原子能科学技术,2014,(3):523-527,5.