Objective To study the accuracy of XE-5000 automatic hematology analyzer in detecting the white blood cell (WBC) classification. Methods 208 cases of outpatients,inpatients and subjects of health examination treated in our hospital from June 2012 to June 2013 were randomly selected and divided into the adult group (100 cases),the children group (60 cases) and the neonate group (48 cases) according to age.The WBC in venous blood of the patients in differ-ent groups were classified respectively by Sysmex XE-5000 method and artificial microscopy method.The results of the two methods were compared and analyzed by using statistical software,and the correlation analysis was carried out. Re-sults In addition to the basophils,the difference of the count results of the WBC classification in the adult group detect-ed by using instrument method and artificial microscopy method was not significant (P>0.05),and had a positive correla-tion (P<0.05).In addition to the monocytes and basophils,the difference of the count results of the WBC classification in the children group and neonate group was not significant (P>0.05),and had a positive correlation (P<0.05). Conclusion Using the Sysmex XE-5000 automatic blood cell analyzer to detect the white blood cell classification has a good per-formance and it is suitable to detect a large quantity of samples quickly and effectively,but it can not completely re-place the artificial microscopy.If the two methods are combined,it can ensure the accuracy and reliability of test results.

关键词

全自动血细胞分析仪/白细胞分类/人工镜检

Key words

Automatic hematology analyzer/White blood cell classification/Artificial microscopy

分类

医药卫生

引用本文复制引用

陈建霞,黄衍锋,张旭..XE-5000全自动血细胞分析仪检测白细胞分类与人工镜检结果对比分析[J].中国当代医药,2015,(12):124-126,3.

中国当代医药

1674-4721

访问量0
|
下载量0
段落导航相关论文