不同检测模式对献血前ALT筛查效果分析
Screening effects analysis of different ALT detection modes prior to blood donation
杜玮璐1
作者信息
- 1. 河南省平顶山市中心血站,河南平顶山 467000
- 折叠
摘要
Abstract
Objective To explore the selection and clinical applicability of alanine aminotransferase (ALT) rapid screening models prior to blood donation. Methods 800 cases of blood donors from May 2013 to August in our central blood bank were detected using rate method and dry chemical method for ALT.The test results of two modes was compared,the two modes was also used to test normal and pathological values quality control serum and their detection precisions was compared. Results The total coincidence rate of dry chemical method was 98.75%.There was no significant difference between the two ALT detection results in the PNU group (P>0.05).In the PPU group,the ALT test results of dry chemical method was higher than that of rate method,with significant difference(P<0.05).The CV results of rate method in two groups was lower than that of dry chemical method,with significant difference(P<0.05). Conclusion Dry chemical assay mode is simple,fast,and reliable,which is suitable for blood donors’ALT screening in street blood collection stations,it is worthy of clinical promotion and application.关键词
无偿献血/干式化学法/速率法/丙氨酸氨基转移酶Key words
Blood donation without payment/Dry chemical method/Rate method/Alanine aminotransferase分类