测量距离影响辐射温度计检定结果的实验研究

栾海峰 王征

计量学报Issue(6):588-590,3.
计量学报Issue(6):588-590,3.DOI:10.3969/j.issn.1000-1158.2014.06.14

测量距离影响辐射温度计检定结果的实验研究

ExperimentaI Research on the Effect of Measuring Distance on CaIibration ResuIts of Radiation Thermometers

栾海峰 1王征1

作者信息

  • 1. 江苏省计量科学研究院,江苏 南京210007
  • 折叠

摘要

Abstract

Aiming at the effect Of measuring distance On caIibratiOn resuIts Of radiatiOn thermOmeters,the experimentaI research was carried Out. BOth testing data and anaIyticaI charts fOr the severaI cOnventiOnaI radiatiOn thermOmeters chOsen are shOwn. The OptimaI measuring distance Of radiatiOn thermOmeter is defined,which is in the range Of 0. 3~0. 5 m tO be away frOm the target surface Of bIackbOdy radiatiOn sOurce.

关键词

计量学/辐射温度计/测量距离/黑体辐射源

Key words

MetrOIOgy/RadiatiOn thermOmeter/Measuring distance/BIackbOdy radiatiOn sOurce