检验医学与临床Issue(13):1792-1793,1795,3.DOI:10.3969/j.issn.1672-9455.2014.13.021
两种全自动血细胞分析仪报警信息功效率的比较
Comparison of the efficiency of flags of two kinds of automatic blood cells analyzer
何廉旗 1马连学 1于欣 1祁立子 1毕淑艳 1张肇乾1
作者信息
- 1. 辽宁省抚顺市抚顺矿务局总医院检验科 113008
- 折叠
摘要
Abstract
Objective To evaluate efficiency of flagging rates of Sysmex XE‐5000 and Sysmex XS‐1000i .Meth‐ods A total of 1 028 whole blood samples were detected by for morphological characteristics by using XE‐5000 ,the XS‐1000i and microscopic method .Results of the former two methods were compared with microscopic method by u‐sing chi‐square analysis .Results Of XS‐1000i ,the blast flag demonstrated 23 false positive ,3 true positive and 1 false negative ,and of XE‐5000 ,there were 5 false positive ,3 true positive and 0 false negative .Of XS‐1000i ,the ab‐normal lymphocyte and atypical lymphocyte flag showed 103 false positive ,30 true positive and 0 false negative ,and of XE‐5000 ,there were 28 false positive ,33 true positive and 0 false negative .Conclusion XE‐5000 might be more ef‐ficient than the XS‐1000i ,which could reduce the review times and improve the efficiency of daily work in laboratory .关键词
XE-5000全自动血细胞分析仪/报警信息/功效率/XS-1000i全自动血细胞分析仪Key words
XE-5000 automatic blood cells analyzer/flags/efficiency/XS-1000i automatic blood cells analyzer引用本文复制引用
何廉旗,马连学,于欣,祁立子,毕淑艳,张肇乾..两种全自动血细胞分析仪报警信息功效率的比较[J].检验医学与临床,2014,(13):1792-1793,1795,3.