摘要

Abstract

Objective To evaluate efficiency of flagging rates of Sysmex XE‐5000 and Sysmex XS‐1000i .Meth‐ods A total of 1 028 whole blood samples were detected by for morphological characteristics by using XE‐5000 ,the XS‐1000i and microscopic method .Results of the former two methods were compared with microscopic method by u‐sing chi‐square analysis .Results Of XS‐1000i ,the blast flag demonstrated 23 false positive ,3 true positive and 1 false negative ,and of XE‐5000 ,there were 5 false positive ,3 true positive and 0 false negative .Of XS‐1000i ,the ab‐normal lymphocyte and atypical lymphocyte flag showed 103 false positive ,30 true positive and 0 false negative ,and of XE‐5000 ,there were 28 false positive ,33 true positive and 0 false negative .Conclusion XE‐5000 might be more ef‐ficient than the XS‐1000i ,which could reduce the review times and improve the efficiency of daily work in laboratory .

关键词

XE-5000全自动血细胞分析仪/报警信息/功效率/XS-1000i全自动血细胞分析仪

Key words

XE-5000 automatic blood cells analyzer/flags/efficiency/XS-1000i automatic blood cells analyzer