| 注册
首页|期刊导航|检验医学与临床|分析和评价强生 Vitros 5.1 FS 生化仪的精密度和正确度

分析和评价强生 Vitros 5.1 FS 生化仪的精密度和正确度

林海标 范雪莲 王建兵 林莉 郑松柏 柯培锋 何敏 万泽民 徐建华

检验医学与临床Issue(15):2047-2048,2050,3.
检验医学与临床Issue(15):2047-2048,2050,3.DOI:10.3969/j.issn.1672-9455.2014.15.003

分析和评价强生 Vitros 5.1 FS 生化仪的精密度和正确度

Performance evaluation of precision and accuracy of Johnson Vitros 5 .1 FS biochemistry analyzer

林海标 1范雪莲 2王建兵 1林莉 1郑松柏 1柯培锋 1何敏 1万泽民 1徐建华1

作者信息

  • 1. 广东省中医院检验科,广州 510120
  • 2. 广东省广州市海珠区妇幼保健院检验科 510240
  • 折叠

摘要

Abstract

Objective To analyze and evaluate the precision and accuracy of Johnson Virtros 5 .1 FS biochem-istry analyzer .Methods According to Clinical and Laboratory Standards Institute (CLSI)guideline(EP15-A2)and other documents ,performance evaluation was made for the precision and accuracy about 17 biochemistry items of Johnson Vitros 5 .1 FS biochemistry analyzer .The results were compared with manufacturer′s claims or acceptable standards .Results The degree of precision displays that Swithin (Stotal)were less than σwithin(σtotal) ,which con-form to the requirements .The reference values were within the allowance range which conform to the requirements of accuracy .Conclusion The precision and accuracy about 17 biochemistry items of Johnson Vitros 5 .1 FS biochemistry analyzer were accorded with quality objectives .

关键词

精密度/正确度/性能评价

Key words

precision/accuracy/performance evaluation

引用本文复制引用

林海标,范雪莲,王建兵,林莉,郑松柏,柯培锋,何敏,万泽民,徐建华..分析和评价强生 Vitros 5.1 FS 生化仪的精密度和正确度[J].检验医学与临床,2014,(15):2047-2048,2050,3.

基金项目

国家中医药行业科研专项基金(201007005);“重大新药创新”国家科技重大发现(2012ZX093030090-300)。 ()