检验医学与临床Issue(z1):53-55,3.DOI:10.3969/j.issn.1672-9455.2014.25.023
应用CLSI EP5-A2文件评价ADAMSTM A1c HA-8160全自动糖化血红蛋白仪的精密度
Evaluation of the Precision Performance of ADAMS HA-8180 HbA1c Analyzer Using CLSI EP5-A2 Guidelines
崔晓阳 1朱彩云 1周卫 1吴敏1
作者信息
- 1. 广东省深圳市龙华人民医院检验科 518109
- 折叠
摘要
Abstract
Objective To evaluate the precision of the Automatic Glycohemoglobin Analyzer ADAMSTM A1c HA-8160.Methods Precision study was performed according to CLSI EP5-A2 guidelines,two(high and normal)lev-els of quality control materials were analyzed.Results For high and normal level quality control materials,the with-run CVs were 0.97% and 1.45%,the between-run CVs were 0.62% and 0.34%,the between-day CVs were 0.63%and 0.56%,and the total CVs were 1.31% and 1.59%,respectively.Conclusion The ADAMSTM A1c HA-8180 HbA1c analyzer showed excellent precision,which could meet the requirements of clinical test.关键词
精密度/糖化血红蛋白/HLPCKey words
precision/glycosylated hemoglobin/high performance liquid chromatography引用本文复制引用
崔晓阳,朱彩云,周卫,吴敏..应用CLSI EP5-A2文件评价ADAMSTM A1c HA-8160全自动糖化血红蛋白仪的精密度[J].检验医学与临床,2014,(z1):53-55,3.