检验医学与临床Issue(14):2072-2073,2.DOI:10.3969/j.issn.1672-9455.2015.14.041
PRIMUS-PDQ Plus 床旁检测仪与 BIO-RAD D10糖化血红蛋白分析仪的一致性评估
Assessment on consistence between PRIMUS-PDQ Plus bedside detector and BIO-RAD D10 analyzers for detecting HbA1 c
张礼强 1李光荣 1常欧 1朱喜丹 1曾章锐 1刘靳波1
作者信息
- 1. 泸州医学院附属医院检验科,四川泸州 646000
- 折叠
摘要
Abstract
Objective To investigate the consistence between PRIMUS-PDQ Plus bedside detector and BIO-RAD D10 analyzer for detecting HbA1c.Methods The method comparison and bias estimation were performed by u-sing patient samples according to the document EP9-A2 of NCCLS.Results The regression equation of HbA1c was Y =0.970 2X +0.274 7 and the correlation coefficient r=0.996 5,the bias of medical decision level was less than the requirement of external quality assessment in the American Society of Clinical Pathologists(ASCP).Conclusion The PRIMUS-PDQ Plus bedside detector and BIO-RAD D10 analyzer possess better consistence.The two equipments have no difference in the detection results,which can meet the goal of clinical diagnosis,treatment and diabetes moni-toring.关键词
糖化血红蛋白仪/床旁检测/偏倚评估/高效液相层析法Key words
glycosylated hemoglobin analyzer/point-of-care testing/bias estimation/high-performance liquid chromatography引用本文复制引用
张礼强,李光荣,常欧,朱喜丹,曾章锐,刘靳波..PRIMUS-PDQ Plus 床旁检测仪与 BIO-RAD D10糖化血红蛋白分析仪的一致性评估[J].检验医学与临床,2015,(14):2072-2073,2.