摘要

Abstract

Objective To investigate the consistence between PRIMUS-PDQ Plus bedside detector and BIO-RAD D10 analyzer for detecting HbA1c.Methods The method comparison and bias estimation were performed by u-sing patient samples according to the document EP9-A2 of NCCLS.Results The regression equation of HbA1c was Y =0.970 2X +0.274 7 and the correlation coefficient r=0.996 5,the bias of medical decision level was less than the requirement of external quality assessment in the American Society of Clinical Pathologists(ASCP).Conclusion The PRIMUS-PDQ Plus bedside detector and BIO-RAD D10 analyzer possess better consistence.The two equipments have no difference in the detection results,which can meet the goal of clinical diagnosis,treatment and diabetes moni-toring.

关键词

糖化血红蛋白仪/床旁检测/偏倚评估/高效液相层析法

Key words

glycosylated hemoglobin analyzer/point-of-care testing/bias estimation/high-performance liquid chromatography

引用本文复制引用

张礼强,李光荣,常欧,朱喜丹,曾章锐,刘靳波..PRIMUS-PDQ Plus 床旁检测仪与 BIO-RAD D10糖化血红蛋白分析仪的一致性评估[J].检验医学与临床,2015,(14):2072-2073,2.