首页|期刊导航|临床眼科杂志|Oculyzer 和 OrbscanⅡ在眼前节测量中的对比研究

Oculyzer 和 OrbscanⅡ在眼前节测量中的对比研究

赵峰 范伟杰 赵贵阳

临床眼科杂志Issue(6):514-516,3.
临床眼科杂志Issue(6):514-516,3.DOI:10.3969/j.issn.1006-8422.2014.06.010

Oculyzer 和 OrbscanⅡ在眼前节测量中的对比研究

Comparison of anterior ocular segment measurements using Oculyzer versus Orbscan II

赵峰 1范伟杰 1赵贵阳1

作者信息

  • 1. 210006 南京医科大学附属南京医院 南京市第一医院 眼科
  • 折叠

摘要

Abstract

Objective To assess the differences and agreements among measurements of eyes using Oculyzer and Orbscan II.The following measurements were conducted:central corneal thickness (CCT), mean keratometry (Km), and anterior chamber depth (ACD).Methods In this prospective clinical study , CCT, Km and ACD were measured in 63 eyes with both Oculyzer and Orbscan II .The results were compared between the devices .Results The mean CCT ob-tained by Oculyzer and Orbscan II were 533.21 ±27.02 μm and 528.18 ±29.12 μm, respectively ( P <0.05).The mean Km were 43.32 ±1.52 D and 42.99 ±1.45 D ( P >0.05), respectively.The mean ACD obtained by Oculyzer and Orbscan II were 2.72 ±0.42 mm and 2.69 ±0.42 mm ( P >0.05).Conclusion Oculyzer and Orbscan II demonstrated a good agreement in measuring CCT , Km and ACD.However,the two devices are not interchangeable for CCT measure-ment .

关键词

眼前节生物测量/角膜中央厚度/角膜曲率/前房深度

Key words

Anterior ocular segment measurements/Central corneal thickness/Mean keratometry/Anterior chamber depth