新鲜抗凝全血对Sysmex不同型号血细胞分析仪比对评价

李鹏 张宁 庄贵华

现代检验医学杂志Issue(6):136-138,3.
现代检验医学杂志Issue(6):136-138,3.DOI:10.3969/j.issn.1671-7414.2014.06.044

新鲜抗凝全血对Sysmex不同型号血细胞分析仪比对评价

Evaluation of Sysmex Hematology Analyzers with Fresh Blood

李鹏 1张宁 1庄贵华2

作者信息

  • 1. 西安交通大学医学院第一附属医院检验科,西安 710061
  • 2. 西安交通大学医学院,西安 710061
  • 折叠

摘要

Abstract

Objective To investigate comparability between Sysmex different hematology analyzers.The consistency of all he-matology analyzers with the same fresh sample.Methods Selected Sysmex XE-2100 as reference instrument which attend external quality assessment.Fresh high,medium and low blood were detected in Sysmex XS-800i,Sysmex XT-4000i and Sys-mex XT-1800i,compared results of WBC,RBC,HGB,HCT,MCV and PLT.Results The bias of XS-800i and XE-2100 was WBC 2.85%,RBC 1.44%,HGB 0.75%,HCT 2.11% and PLT 5.53%.The bias of XT-4000i and XE-2100 was WBC 1.26%,RBC 0.95%,HGB 0.68%,HCT 1.35% and PLT 2.68%.The bias of XT-1800i and XE-2100 was WBC 5.21%, RBC 1.96%,HGB 1.60%,HCT 1.96% and PLT 4.95%.It had good compatability.All various parameters was in the al-lowed range.Conclusion Should maintenance and compare the hematology analyzers periodically,found the problem then calibration instruments in time,ensure the consistency of measurement results between different instruments and guarantee accuracy of the results.

关键词

血细胞分析仪/比对试验/质量控制

Key words

hematology analyzers/comparative test/quality control