中国医疗设备Issue(4):120-122,3.DOI:10.3969/j.issn.1674-1633.2014.04.046
无创血压监护仪在心房纤颤病人血压测量中的准确性评价
Accuracy Evaluation of Non-invasive Blood Pressure Monitor in Blood Pressure Measurement of Patients with Atrial Fibrillation
胡文志 1周海波1
作者信息
- 1. 南京医科大学第二附属医院心内科,江苏南京210011
- 折叠
摘要
Abstract
Objective To investigate the accuracy of non-invasive blood pressure monitor in blood pressure measurement of patients with atrial fibrillation. Methods The blood pressure of 60 patients with atrial fibrillation was measured by using non-invasive blood pressure monitor and mercury sphygmomanometer, respectively. Then the measuring results of blood pressure of non-invasive blood pressure monitor were compared with those of mercury sphygmomanometer. Results There was no signiifcant difference between the measuring results of blood pressure of non-invasive blood pressure and those of mercury sphygmomanometer on systolic pressure (P>0.05) while there were signiifcant differences between the measuring results of blood pressure of non-invasive blood pressure and those of mercury sphygmomanometer on diastolic pressure (P<0.05). Conclusion It is inaccurate to measure the blood pressure of patients with atrial ifbrillation by using non-invasive blood pressure monitor, which should be paid attention to in clinic.关键词
无创血压监护仪/汞柱台式血压计/心房纤颤/收缩压/舒张压Key words
non-invasive blood pressure monitor/mercury sphygmomanometer/atrial fibrillation/systolic pressure/diastolic pressure分类
医药卫生引用本文复制引用
胡文志,周海波..无创血压监护仪在心房纤颤病人血压测量中的准确性评价[J].中国医疗设备,2014,(4):120-122,3.