无创血压监护仪在心房纤颤病人血压测量中的准确性评价

Accuracy Evaluation of Non-invasive Blood Pressure Monitor in Blood Pressure Measurement of Patients with Atrial Fibrillation

胡文志 1周海波1

作者信息

  • 1. 南京医科大学第二附属医院心内科,江苏南京210011
  • 折叠

摘要

Abstract

Objective To investigate the accuracy of non-invasive blood pressure monitor in blood pressure measurement of patients with atrial fibrillation. Methods The blood pressure of 60 patients with atrial fibrillation was measured by using non-invasive blood pressure monitor and mercury sphygmomanometer, respectively. Then the measuring results of blood pressure of non-invasive blood pressure monitor were compared with those of mercury sphygmomanometer. Results There was no signiifcant difference between the measuring results of blood pressure of non-invasive blood pressure and those of mercury sphygmomanometer on systolic pressure (P>0.05) while there were signiifcant differences between the measuring results of blood pressure of non-invasive blood pressure and those of mercury sphygmomanometer on diastolic pressure (P<0.05). Conclusion It is inaccurate to measure the blood pressure of patients with atrial ifbrillation by using non-invasive blood pressure monitor, which should be paid attention to in clinic.

关键词

无创血压监护仪/汞柱台式血压计/心房纤颤/收缩压/舒张压

Key words

non-invasive blood pressure monitor/mercury sphygmomanometer/atrial fibrillation/systolic pressure/diastolic pressure

分类

医药卫生

引用本文复制引用

胡文志,周海波..无创血压监护仪在心房纤颤病人血压测量中的准确性评价[J].中国医疗设备,2014,(4):120-122,3.

中国医疗设备

OACSTPCD

1674-1633

访问量0
|
下载量0
段落导航相关论文