中国医疗设备Issue(4):123-124,2.DOI:10.3969/j.issn.1674-1633.2014.04.047
脉动真空灭菌器B-D试验问题分析
Problem Analysis of B-D Test of Pulsating Vacuum Sterilizer
南光1
作者信息
- 1. 吉林市人民医院设备科,吉林长春130021
- 折叠
摘要
Abstract
This paper summarized and analyzed the failure phenomena with their causes of B-D test during the sterilization process of pulsating vacuum sterilizer. Then the pulsating vacuum sterilizer was remoulded according to the causes of failure phenomena of B-D test. Thus the problem of B-D test failure was solved.关键词
脉动真空灭菌器/B-D试验/冷凝水/疏水器Key words
pulsating vacuum sterilizer/B-D test/condensate water/steam trap分类
医药卫生引用本文复制引用
南光..脉动真空灭菌器B-D试验问题分析[J].中国医疗设备,2014,(4):123-124,2.