手术室C形臂X射线机的放射防护检测与对比研究

Comparative Study and Detection of C-arm X-ray Machine Radiation Protection in the Operation Room

高华永 1刘帅 1苗路瑞 1蒋建 1晁勇1

作者信息

  • 1. 中国人民解放军总医院第一附属医院医学工程科,北京 100048
  • 折叠

摘要

Abstract

The X-ray radiation doses in protected and non-protected operation room were measured. By analyzing the measurement data, it is proved that C-arm X-ray machine, without protection measures, will produce radiation hazard to the staff around the operation room. The solutions to the prevention of ionizing radiation in non-protected operation room are discussed in this paper.

关键词

C形臂X射线机/放射防护/手术间

Key words

C-arm X-ray machine/radiation protection/operation room