Monaco与Xio治疗计划系统调强放疗剂量验证的比较研究
Comparative Study on the Dosimetric Veriifcation of Intensity-Modulated Radiation Therapy of Monaco and Xio Treatment Planning Systems
王宁 1陈阿龙 2夏景涛3
作者信息
- 1. 中山市中医院 放射科,广东 中山 528400
- 2. 中山大学肿瘤防治中心 放疗科,广东 广州 510060
- 3. 南方医科大学南方医院,广东 广州 510515
- 折叠
摘要
Abstract
Objective To evaluate the clinical accuracy of Monaco and Xio treatment planning systems (TPS) by comparing the dose verification results of intensity-modulated radiation therapy (IMRT) of the two treatment planning systems.Methods The dose veriifcation of IMRT plans for nasopharyngeal carcinoma formulated by Xio TPS (n=20) and Monaco TPS (n=20) was conducted by IBA Matrixx. Then Gamma pass rate analysis was conducted according to various criterias including 2%/2 mm, 3%/3 mm and 5%/3 mm.Results The average Gamma pass rate of IMRT plans of Monaco TPS was higher than that of Xio TPS according to all the criterias. There were signiifcant differences between Monaco TPS and Xio TPS on the Gamma pass rate according to 2%/2 mm and 3%/3 mm criterias (P<0.05) while there were no signiifcant differences between the two treatment planning systems on the Gamma pass rate according to 5%/3 mm criterion (P>0.05).Conclusion Both of Monaco and Xio treatment planning systems can be applied to clinical IMRT while the accuracy of Monaco TPS is higher than that of Xio TPS.关键词
Xio治疗计划系统/Monaco治疗计划系统/调强放疗/剂量验证/Gamma通过率Key words
Xio TPS/Monaco TPS/IMRT/dose verification/Gamma pass rate分类
医药卫生引用本文复制引用
王宁,陈阿龙,夏景涛..Monaco与Xio治疗计划系统调强放疗剂量验证的比较研究[J].中国医疗设备,2014,(7):20-22,3.