中国医疗设备Issue(7):53-54,22,3.DOI:10.3969/j.issn.1674-1633.2014.07.019
2012~2013年内蒙古自治区医用电子直线加速器性能检测结果分析
Analysis on Performance Detection Results of Medical Linear Accelerators in Inner Mongolia From 2012 to 2013
摘要
Abstract
Objective To summarize and analyze the performance detection results of medical linear accelerators in Inner Mongolia from 2012 and 2013 to discuss the change trends with corresponding causes of the detection rate and qualiifed rate of the accelerators.Methods The performance detection of the accelerators was conducted through the methods proposed in national occupational health standards GB15213-94, GB/T19046-2003 and GBZ126-2011.Results The detection rate and qualified rate of the accelerators in 2012 were 44% and 72.7%, respectively. The detection rate and qualiifed rate in the initial detection of the accelerators in 2013 were 92.6% and 92%, respectively. The qualiifed rate of the accelerators in 2013 was raised to 96% by repairing several unqualiifed accelerators.Conclusion The detection rate and qualiifed rate of the accelerators had been signiifcantly improved in Inner Mongolia through the equipment replacement, the active equipment detection by testing agencies and the equipment maintenance by radiotherapy institutions.关键词
医用电子直线加速器/X射线检测/电子线检测/检测率/合格率Key words
medical linear accelerator/X-ray detection/electron beam detection/detection rate/qualiifed rate分类
医药卫生引用本文复制引用
哈日巴拉,张淼,许潇,格日勒满达呼,王成国..2012~2013年内蒙古自治区医用电子直线加速器性能检测结果分析[J].中国医疗设备,2014,(7):53-54,22,3.基金项目
卫生行业科研专项(201002009)。 ()