| 注册
首页|期刊导航|中国医疗设备|2012~2013年内蒙古自治区医用电子直线加速器性能检测结果分析

2012~2013年内蒙古自治区医用电子直线加速器性能检测结果分析

哈日巴拉 张淼 许潇 格日勒满达呼 王成国

中国医疗设备Issue(7):53-54,22,3.
中国医疗设备Issue(7):53-54,22,3.DOI:10.3969/j.issn.1674-1633.2014.07.019

2012~2013年内蒙古自治区医用电子直线加速器性能检测结果分析

Analysis on Performance Detection Results of Medical Linear Accelerators in Inner Mongolia From 2012 to 2013

哈日巴拉 1张淼 2许潇 1格日勒满达呼 1王成国1

作者信息

  • 1. 内蒙古疾病预防控制中心,内蒙古 呼和浩特 010031
  • 2. 内蒙古自治区人民医院 医学工程处,内蒙古 呼和浩特 010017
  • 折叠

摘要

Abstract

Objective To summarize and analyze the performance detection results of medical linear accelerators in Inner Mongolia from 2012 and 2013 to discuss the change trends with corresponding causes of the detection rate and qualiifed rate of the accelerators.Methods The performance detection of the accelerators was conducted through the methods proposed in national occupational health standards GB15213-94, GB/T19046-2003 and GBZ126-2011.Results The detection rate and qualified rate of the accelerators in 2012 were 44% and 72.7%, respectively. The detection rate and qualiifed rate in the initial detection of the accelerators in 2013 were 92.6% and 92%, respectively. The qualiifed rate of the accelerators in 2013 was raised to 96% by repairing several unqualiifed accelerators.Conclusion The detection rate and qualiifed rate of the accelerators had been signiifcantly improved in Inner Mongolia through the equipment replacement, the active equipment detection by testing agencies and the equipment maintenance by radiotherapy institutions.

关键词

医用电子直线加速器/X射线检测/电子线检测/检测率/合格率

Key words

medical linear accelerator/X-ray detection/electron beam detection/detection rate/qualiifed rate

分类

医药卫生

引用本文复制引用

哈日巴拉,张淼,许潇,格日勒满达呼,王成国..2012~2013年内蒙古自治区医用电子直线加速器性能检测结果分析[J].中国医疗设备,2014,(7):53-54,22,3.

基金项目

卫生行业科研专项(201002009)。 ()

中国医疗设备

OACSTPCD

1674-1633

访问量0
|
下载量0
段落导航相关论文