气腹机性能检测试验装置的研究OACSTPCD
Research on Testing Devices for Performance Detection of Insuflfator
本文分析了行业标准中对气腹机气压和耗气量测定的要求,并按照要求设计了患者腹腔模拟系统以及耗气量测定容器,并应用两种装置对两款气腹机的性能进行了测试,结果表明设计的两种装置能够应用于气腹机性能检测。
This paper analyzed the requirements of industry standard on insufflators in terms of air pressure and gas consumption measurements. According to the requirements of industry standard, the analog system of patients’ abdominal cavity and the container for gas consumption measurement were designed to detect the performance of two kinds of insuflfators. The detection results showed that the testing devices can be used in the performance detection of insuflfators.
王晨希;刘艳珍;王权;孟祥峰;任海萍
中国食品药品检定研究院,北京 100050中国食品药品检定研究院,北京 100050中国食品药品检定研究院,北京 100050中国食品药品检定研究院,北京 100050中国食品药品检定研究院,北京 100050
机械制造
气腹机模拟腹腔气压测定耗气量测定
insufflatoranalog abdominal cavityair pressure measurementgas consumption measurement
《中国医疗设备》 2014 (8)
112-114,3