中国医疗设备Issue(9):53-55,3.DOI:10.3969/j.issn.1674-1633.2014.09.018
数字减影血管造影系统性能检测和质量控制
Performance Testing and Quality Control of Digital Subtraction Angiography System
胡益斌 1扈尚泽2
作者信息
- 1. 江苏省人民医院 医学工程处,江苏南京 210029
- 2. 江苏省计量科学研究院,江苏南京210007
- 折叠
摘要
Abstract
In order to ensure the safety and effectiveness of digital subtraction angiography (DSA) system, it is essential to conducted performance testing and control quality of DAS system regularly. According to the national metrology veriifcation regulation (JJG 1067-2011) of X-ray radiation source of medical diagnostic digital subtraction angiography (DSA) system, this paper introduced the testing methods of air kerma rate, HVL, X-ray tube voltage, minimal size of simulate vascular, spatial resolution, low contrast resolution, contrast linearity and subtraction angiography performance impact.关键词
数字减影血管造影系统/空气比释动能率/X射线管电压/计量检定/性能检测Key words
DSA system/air kerma rate/X-ray tube voltage/metrological verification/performance testing分类
机械制造引用本文复制引用
胡益斌,扈尚泽..数字减影血管造影系统性能检测和质量控制[J].中国医疗设备,2014,(9):53-55,3.