| 注册
首页|期刊导航|中国医疗设备|OmniPro I’mRT分析软件与平板式扫描仪在螺旋断层放疗计划验证中的应用

OmniPro I’mRT分析软件与平板式扫描仪在螺旋断层放疗计划验证中的应用

刘海 杭霞瑜 李益坤 张新良 丁巍

中国医疗设备Issue(12):112-114,118,4.
中国医疗设备Issue(12):112-114,118,4.DOI:10.3969/j.issn.1674-1633.2014.12.037

OmniPro I’mRT分析软件与平板式扫描仪在螺旋断层放疗计划验证中的应用

Applicafion of OmniPro I’mRT Combined with Flatbed RGB Scanner in the Verification of Helical Tomotherapy Plans

刘海 1杭霞瑜 1李益坤 1张新良 1丁巍1

作者信息

  • 1. 中国人民解放军第八一医院放疗科,江苏 南京 210002
  • 折叠

摘要

Abstract

Objective To investigate the application value of OmniPro I’mRT (OI)+RGB fiatbed scanner (FS) in the verification of helical tomotherapy (HT) plans.Methods 12 HT plans were randomly selected, and GAFchromic EBT3 films placed in CheesePhantom were irradiated to obtain the coronal plane dose distribution. Then the γanalysis was used in comparing and analyzing the verification effects of OI+FS program and the tomo planning station (TomoPS) DQA analysis software+ Vidar film digitizer (VidarFD) program.ResuItsThe average pass rate of TomoPS+VidarFD program was (97.32%±1.16%), while that of OI+FS program was (96.40%±1.01 %), which indicated that the average pass rates of the two programs were high. The Pearson correlation analysis results showed that the verification data of OI+FS program was closely related to that of TomoPS+VidarFD program (r=0.919,P<0.05). ConcIusionOmniPro I ‘mRT analysis software combined with RGB fiated scanner can be used as a method in verifying the HT plans.

关键词

螺旋断层放射治疗/剂量验证/γ分析/胶片数字化仪/质量保证

Key words

helical tomotherapy/dosimetric verification/γ analysis/film digitizer/quality assurance

分类

机械制造

引用本文复制引用

刘海,杭霞瑜,李益坤,张新良,丁巍..OmniPro I’mRT分析软件与平板式扫描仪在螺旋断层放疗计划验证中的应用[J].中国医疗设备,2014,(12):112-114,118,4.

中国医疗设备

OACSTPCD

1674-1633

访问量0
|
下载量0
段落导航相关论文