心律失常仿真下监护仪无创血压测量误差的Loess回归分析

王晓龙 白玫

中国医疗设备Issue(5):30-32,3.
中国医疗设备Issue(5):30-32,3.DOI:10.3969/j.issn.1674-1633.2015.05.009

心律失常仿真下监护仪无创血压测量误差的Loess回归分析

Analysis of the Patient Monitors’ NIBP Measurement Errors in the Case of Arrhythmia Simulations through Loess Regression

王晓龙 1白玫1

作者信息

  • 1. 首都医科大学宣武医院,北京 100053
  • 折叠

摘要

Abstract

Objective To discuss the inlfuence of arrhythmia on NIBP (Non-Invasive Blood Pressure) measurement of the patient monitor. Methods FLUKE ProSim8 Vital Signs Simulator was deployed to simulate NSR (Normal Sinus Rhythm) and ifve kinds of arrhythmia signals, including VT (Ventricular Tachycardia), SB (Sinus Bradycardia), PT (Paroxysmal Tachycardia), PVC (Premature Ventricular Contraction) and AT (Auricular Tachycardia). Meanwhile, 4 groups of NIBP measurement values were given under each kind of simulation signals:60/30, 120/80, 150/110, 200/150 (mmHg). And a total of 30 GE Pro1000 and Dash2500 patient monitors were selected to measure the given blood pressure and calculate the errors. Then, Loess (Locally-Weighted Scatterplot Smoothing) regression was adopted to ift the errors. Results The curves worked out by Loess regression highlighted that the curve line of VT-NSR was obviously higher than that of the normal group. Conclusion An extremely high heart rate might cause a higher NIBP measurement result of the patient monitor. However, a low heart rate caused no obvious inlfuence on the measurement result. Therefore, the patient monitors can not detect the NIBP accurately when a patient suffered from an extremely high heart rate.

关键词

心律失常/无创血压测量/监护仪/生命体征模拟器

Key words

arrhythmia/noninvasive blood pressure measurement/patient monitor/vital signs simulator

分类

医药卫生

引用本文复制引用

王晓龙,白玫..心律失常仿真下监护仪无创血压测量误差的Loess回归分析[J].中国医疗设备,2015,(5):30-32,3.

中国医疗设备

OACSTPCD

1674-1633

访问量0
|
下载量0
段落导航相关论文