原子能科学技术Issue(6):974-979,6.DOI:10.7538/yzk.2014.48.06.0974
1AP 中铀钚在线分析
On-line Determination of Uranium and Plutonium in 1AP
康海英 1郑维明 1吴继宗 1宋游 1陈晨 1刘桂娇1
作者信息
- 1. 中国原子能科学研究院 放射化学研究所,北京 102413
- 折叠
摘要
Abstract
In order to determine the concentration of uranium and plutonium in 1AP of bench experiment in time ,an on-line X-ray fluorescence spectrometer was designed and set up ,and the angle between incident X-ray and emitted X-ray was 60° .The spectrom-eter was airproofed in order to protect the X-ray tube and the detector .Based on the plotted working curve of uranium and plutonium ,the analytical method was established to determine the concentration of uranium and plutonium in 1AP .The relative standard deviation of the measured concentration of uranium is 0.42% . T he relative range for repeated measurement of uranium concentration in three hours is 1.6% . T he results show that the device is stable enough .This spectrometer was applied in on-line determi-nation of uranium .It can monitor the fluctuation of uranium concentration ,and the results agree well with that of titration means .关键词
铀/钚/X射线荧光/1AP/在线测量Key words
uranium/plutonium/X-ray fluorescence/1AP/on-line determination分类
化学化工引用本文复制引用
康海英,郑维明,吴继宗,宋游,陈晨,刘桂娇..1AP 中铀钚在线分析[J].原子能科学技术,2014,(6):974-979,6.