原子能科学技术Issue(3):572-576,5.DOI:10.7538/yzk.2015.49.03.0572
铅铋合金中Bi/Bi2O3型氧传感器准确性及稳定性测试研究
Accuracy and Stability Test Study of Bi/Bi2 O3 Oxygen Sensor in Lead-bismuth Eutectic
摘要
Abstract
Accurate measurement of the concentration of dissolved oxygen in liquid lead‐bismuth eutectic (LBE) is the premise to achieve oxygen control .A self‐developed Bi/Bi2 O3 oxygen sensor was used to measure oxygen concentration in LBE .The accuracy and stability tests were carried out in stagnant liquid LBE in the condition of oxygen sat‐uration .The accuracy test results show that the change of experimental electromotive force with temperature is consistent with the theoretical curve in the temperature ranging from 340 ℃ to 480 ℃ .The maximum absolute error of electromotive force is-2.6 mV .The stability test results show that the absolute error of electromotive force is -1.4 mV within 50 hours at the temperature of 450 ℃ and the fluctuation in this test is4mV .关键词
铅铋合金/氧浓度/氧传感器/电动势Key words
lead-bismuth eutectic/oxygen concentration/oxygen sensor/electromotive force分类
能源科技引用本文复制引用
王艳青,黄群英,武欣,吴斌,张敏,高胜..铅铋合金中Bi/Bi2O3型氧传感器准确性及稳定性测试研究[J].原子能科学技术,2015,(3):572-576,5.基金项目