中国口腔种植学杂志Issue(2):62-65,4.
个性化拟自然牙种植体的设计及有限元分析
Design and finite element analysis of custom-made, root-analogue implant
高亦林 1游嘉 1彭伟 1周见 2田明飞2
作者信息
- 1. 310014 浙江工业大学大学特种装备制造与先进加工技术教育部重点实验室
- 2. 324000 衢州市人民医院
- 折叠
摘要
Abstract
Objective: To design a kind of custom-made, root-analogue implant for immediate implant placement. Propose the design requirements of the interference of the implant by finite element analysis. Methods: Design and manufacture the implant with the help of RE and 3D printing technology. Design 10 custom-made, root-analogue implants with the interference of 10, 20, 30, 40, 50, 60, 70, 80, 90, 100μm respectively. Simulate the implantation with the same extraction socket by finite element analysis. Results: Under the same condition, stress of cortical bone area is larger than cancellous bone area. With the increasing amount of interference, the greater the stress of implant-bone interface and cortical bone area reach the yield strength first. Conclusions: In the process of designing a custom-made root-analogue implant, the amount of interference of cortical and cancellous bone area should be designed differently. And the amount of interference of cortical bone area should be less than cancellous bone area's.关键词
个性化/拟自然牙/种植体/逆向工程/3D打印/过盈量Key words
custom-made/root-analogue/implant/RE/3D print/interference分类
医药卫生引用本文复制引用
高亦林,游嘉,彭伟,周见,田明飞..个性化拟自然牙种植体的设计及有限元分析[J].中国口腔种植学杂志,2015,(2):62-65,4.