| 注册
首页|期刊导航|中国医学创新|AVE-763B型自动尿沉渣分析仪检测红细胞影响因素的分析

AVE-763B型自动尿沉渣分析仪检测红细胞影响因素的分析

张丽丽

中国医学创新Issue(25):130-132,3.
中国医学创新Issue(25):130-132,3.DOI:10.3969/j.issn.1674-4985.2014.25.045

AVE-763B型自动尿沉渣分析仪检测红细胞影响因素的分析

Analysis on Influencing Factors Regarding Determination of Urine Red Blood Cells by AVE-763B Automated Urinalysis Analyzer

张丽丽1

作者信息

  • 1. 四川省资阳市雁江区人民医院 四川 资阳 641300
  • 折叠

摘要

Abstract

Objective:To analyze the influencing factors regarding determination of urine red blood cells by AVE-763B automated urinalysis analyzer.Method: 545 samples of urine from hospitalized patients were tested by AVE-763B automated urinalysis analyzer and microscopic examination. The results were compared with statistics method.Result: The false-positive rate, false-negative rate, sensitivity and specificity of determination of urine red blood cells by AVE-763B automated urinalysis analyzer were 15.2%, 0,1%, 84.8% respectively. When compared with the microscopic examination group, the results were statistic difference (P<0.05). The results showed that the crystal, epithelial cells, yeast were main factors of affecting the detected results by AVE-763B automated urinalysis analyzer.Conclusion: Discussion AVE-763B automated urinalysis analyzer must combine microscope examination in order to improve the urine examination quality.

关键词

尿沉渣/红细胞/尿沉渣分析仪

Key words

Urinalysis/Erythrocytes/Urinalysis analyzer

引用本文复制引用

张丽丽..AVE-763B型自动尿沉渣分析仪检测红细胞影响因素的分析[J].中国医学创新,2014,(25):130-132,3.