| 注册
首页|期刊导航|中国医学装备|剂量分布仪对医用电子直线加速器X射线泄漏辐射检测方法的探讨

剂量分布仪对医用电子直线加速器X射线泄漏辐射检测方法的探讨

杜翔 曹兴江 杨春勇 王进

中国医学装备Issue(7):4-6,3.
中国医学装备Issue(7):4-6,3.DOI:10.3969/J.ISSN.1672-8270.2014.07.002

剂量分布仪对医用电子直线加速器X射线泄漏辐射检测方法的探讨

The exploration of the measurement methods of the leakage dose level in medical accelerators by using dose scanners

杜翔 1曹兴江 1杨春勇 1王进1

作者信息

  • 1. 江苏省疾病预防控制中心放射防护所 江苏 南京 210009
  • 折叠

摘要

Abstract

Objective: To measure and analysis the leakage dose of a medical accelerator and explore the method for the leakage dose level measurements of the accelerators for medical uses. Methods:A Startrack model QA detector was used to measure the leakage dose of a medical accelerator by using the methods in GBZ126-2011. Results:The maximum leakage dose in M area to maximum dose ratio is 0.07%and the average leakage dose in M area to maximum dose ratio is 0.04%. Conclusion: It is necessary to explore a method to measure the leakage dose level of medical accelerators by dose scanners.

关键词

泄漏辐射/放射防护/医用加速器

Key words

Leakage dose/Medical accelerator/Radiation protection

分类

医药卫生

引用本文复制引用

杜翔,曹兴江,杨春勇,王进..剂量分布仪对医用电子直线加速器X射线泄漏辐射检测方法的探讨[J].中国医学装备,2014,(7):4-6,3.

基金项目

卫生行业科研专项(201002009)“辐射危害控制与核辐射卫生应急处置关键技术研究及其应用”,江苏省“十二·五”科教兴卫工程(ZX201109)“突发公共卫生事件应急处置创新平台” (201002009)

中国医学装备

OACSTPCD

1672-8270

访问量0
|