中国医学装备Issue(3):10-11,12,3.DOI:10.3969/J.ISSN.1672-8270.2015.03.004
对广西15家医院CR系统质量控制检测结果的分析
Research on the test results and analysis for the quality control of CR in Guangxi
摘要
Abstract
Objective:To understand the qualification of CR system in Guangxi, in order to take corresponding quality control strategy.Methods: Select 15 sets of CR system for detection and the research object, according to the implementation plan of 2010 annual health industry research to determine the performance index, and on the basis of national standards on the performance of CR system for testing and analysis.Results: The qualified rate of 15 sets CR system is only 20%, CR special detecting project main unqualified index have IP response uniformity and consistency, IP response linear, IP erasing completeness, etc.Conclusion: To strengthen the quality control of CR system, to maintain the relevant technical indexes on a regular basis, it is necessary to improve the quality of medical image diagnosis.关键词
CR系统/质量控制/设备性能/维护保养Key words
CR system/Quality control/Equipment performance/Maintenance分类
医药卫生引用本文复制引用
陈掌凡,谢萍,董颖,赵新春,梁挺,张会敏..对广西15家医院CR系统质量控制检测结果的分析[J].中国医学装备,2015,(3):10-11,12,3.基金项目
卫生行业科研专项(201002009)“辐射危害控制与核辐射卫生应急处置关键技术研究及其应用”,广西自然科学基金(2014GXNSFAA118223)“广西医用辐射危害评价与控制技术研究”。 (201002009)