| 注册
首页|期刊导航|中国医学装备|192Ir后装治疗机辐射防护及性能检测的评价

192Ir后装治疗机辐射防护及性能检测的评价

林珠 陈秀如 张武哲

中国医学装备Issue(6):31-33,3.
中国医学装备Issue(6):31-33,3.DOI:10.3969/J.ISSN.1672-8270.2015.06.010

192Ir后装治疗机辐射防护及性能检测的评价

Evaluation on radiation protection and performance test for 192Ir brachytherapy apparatus

林珠 1陈秀如 1张武哲1

作者信息

  • 1. 汕头大学医学院附属肿瘤医院放疗科 广东 汕头 515031
  • 折叠

摘要

Abstract

Objective: To evaluate the radiation protection and performance test for the 192Ir brachytherapy apparatus. Methods: According to the national standard, radiation protection and its various properties of the 192Ir brachytherapy apparatus were tested. The radiation leakage dose of the 192Ir brachytherapy apparatus and the radiation scattered dose of the surrounding environment were tested with the ion chamber survey meter 451P. The activity of 192Ir radiation source was calibrated with the well chamber HDR1000 Plus. Results: The radiation leakage and scattered dose of the 192Ir source storage device and the surrounding environment are far below the national standard. The deviation of the activity calibration for 192Ir source is-2.07%, accordance with national standard±5%. Conclusion: Radiation protection and its various properties of the 192Ir brachytherapy apparatus should be regularly monitor for assuring the quality of the brachytherapy.

关键词

后装放射治疗/辐射防护/性能检测

Key words

Brachytherapy/Radiation protection/Performance test

分类

医药卫生

引用本文复制引用

林珠,陈秀如,张武哲..192Ir后装治疗机辐射防护及性能检测的评价[J].中国医学装备,2015,(6):31-33,3.