Analysis on hospital medical equipment electrical safety testing

张华伟 1秦航 1蒋红兵1

作者信息

  • 1. 南京医科大学附属南京医院医疗设备处 江苏 南京 210006
  • 折叠

摘要

Abstract

Objective:To study the electrical safety of multi-parameter monitor and defibrillation monitor, and guarantee the safe and effective operation by through the medical equipment electrical safety performance in our hospital.Methods: To test 101 different models of multi-parameter monitor and defibrillation monitor electrical safety.Results: The one-time pass rate of 96%, and the two-time rate was 98%, also one multi-parameter monitor and defibrillation monitor hadn’t meet electrical safety performance after repairing.Conclusion: The equipments that don’t meet electrical safety performance prohibit using for clinical purposes; To ensure the safety of medical equipment and life safety of medical workers and patients by through developing the testing of the medical equipment electrical safety performance in our hospital.

关键词

医疗设备/电气安全/接地阻抗/绝缘阻抗/漏电流

Key words

Medical equipment/Electrical safety/Grounding resistance/Earth resistance/Leakage current