医院医疗设备电气安全检测分析
Analysis on hospital medical equipment electrical safety testing
张华伟 1秦航 1蒋红兵1
作者信息
- 1. 南京医科大学附属南京医院医疗设备处 江苏 南京 210006
- 折叠
摘要
Abstract
Objective:To study the electrical safety of multi-parameter monitor and defibrillation monitor, and guarantee the safe and effective operation by through the medical equipment electrical safety performance in our hospital.Methods: To test 101 different models of multi-parameter monitor and defibrillation monitor electrical safety.Results: The one-time pass rate of 96%, and the two-time rate was 98%, also one multi-parameter monitor and defibrillation monitor hadn’t meet electrical safety performance after repairing.Conclusion: The equipments that don’t meet electrical safety performance prohibit using for clinical purposes; To ensure the safety of medical equipment and life safety of medical workers and patients by through developing the testing of the medical equipment electrical safety performance in our hospital.关键词
医疗设备/电气安全/接地阻抗/绝缘阻抗/漏电流Key words
Medical equipment/Electrical safety/Grounding resistance/Earth resistance/Leakage current分类
医药卫生引用本文复制引用
张华伟,秦航,蒋红兵..医院医疗设备电气安全检测分析[J].中国医学装备,2015,(7):25-26,27,3.