对比剂浓度与DynaCT影像质量的相关性研究

黄奋尧 俞家熙 陈智杰 方鸣 宋玉全

中国医药科学Issue(9):162-164,3.
中国医药科学Issue(9):162-164,3.

对比剂浓度与DynaCT影像质量的相关性研究

Correlation of the contrast medium concentration and DynaCT image quality

黄奋尧 1俞家熙 1陈智杰 1方鸣 1宋玉全1

作者信息

  • 1. 广州医科大学附属第一医院放射科,广东广州 510120
  • 折叠

摘要

Abstract

Objective To investigate the effects of contrast medium concentration on DynaCT image quality. Methods Group 1: Fill the tubes(d=3mm) with different concentrations of contrast mediums, then photograph with DynaCT. Evaluate the interfere degree of different concentration of contrast mediums to the image quality. Group 2:Choose different diameter of the plastic pipe, respectively. Filled the concentration of 200mg/mL, 100mg/mL contrast medium in the tube, and divided into two groups to evaluate the interfere degree of different inner diameter. Results (1)The greater contrast medium concentration is, the more serious interference is. When the contrast medium is at a concentration of 100mg/mL, the interference disappeared. (2)When the contrast medium concentration is certain, the larger the diameter is, the more severe the interference is. Conclusion In the DynaCT photography, different concentration of contrast medium should be selected based on the size of the target vessel diameter in order to reduce the effect of the contrast medium on image quality. When most of the target vessels are less than 3mm, the contrast medium concentration should better be less than 100mg/mL.

关键词

对比剂浓度/DynaCT

Key words

Contrast medium concentration/DynaCT

分类

医药卫生

引用本文复制引用

黄奋尧,俞家熙,陈智杰,方鸣,宋玉全..对比剂浓度与DynaCT影像质量的相关性研究[J].中国医药科学,2015,(9):162-164,3.

中国医药科学

2095-0616

访问量0
|
下载量0
段落导航相关论文