Abstract

Objective To explore the relationship and difference of serum K+ 、Na+ 、Cl- of different detection systems .Methods Choose high K+ serum was diluted serially ,determination of K+ 、Na+ 、Cl- in SEMENS ADVIA 2400 biochemical analyzer and Backman LX‐20 biochemical analyzer ,the data were analyzed statistically related .Re‐sults When serum K+ results is greater than 7 mmol/L ,two detection system deviation is greater than the clinical laboratory improvement specification CLIA′88 of the maximum allowable total error .Serum Na+ 、Cl-l results be‐tween the two detection system deviation less than CLIA′88 of the range of the maximum allowed .They are notable relationship between the two detection system .Conclusion Determination of serum electrolytes especially K+ must be clear of the detection system can report scope ,and do the level of quality control of the detection system .

关键词

检测系统/相关性/差异

Key words

detection system/relationship/difference

引用本文复制引用

王友祥,尹向飞,梁小亮..血清K+、Na+、Cl-在两个检测系统测定的相关性及差异分析[J].检验医学与临床,2014,(z2):232-234,3.