| 注册
首页|期刊导航|原子能科学技术|溶液中钚α活度的直接测量研究

溶液中钚α活度的直接测量研究

刘联伟 吴继宗 刘权卫 谈树苹

原子能科学技术Issue(8):1467-1472,6.
原子能科学技术Issue(8):1467-1472,6.DOI:10.7538/yzk.2015.49.08.1467

溶液中钚α活度的直接测量研究

Study on Direct Measurement of Plutonium αActivity in Solution

刘联伟 1吴继宗 1刘权卫 1谈树苹1

作者信息

  • 1. 中国原子能科学研究院放射化学研究所,北京 102413
  • 折叠

摘要

Abstract

The YAP∶Ce scintillator was selected as a core component for the instru‐ment , and a contact‐type measurement of micro and trace plutonium α activity in solution was established . The results show that under the aqueous phase plutonium solution concentration from 5.20 × 10-5 g/L to 1.30 × 10-3 g/L ,the concentration of plutonium and net count rate have fine linearity (R2 = 0.987 7) ,and the quantitative detection limit is 5.20 × 10 -5 g/L .Under the organic phase plutonium solution concen‐tration from 2.27 × 10-5 g/L to 1.13 × 10-3 g/L ,the net count rate and the concentra‐tion of plutonium have fine linearity (R2 =0.992 3) ,and the quantitative detection limit is 2.27 × 10 -5 g/L .The method is different from the traditional α‐counting method ,and has the advantages of simple and convenient operation and without complicated sample preparation process .It is expected to provide a new way for micro and trace plutonium determination online or in laboratory in spent fuel reprocessing .

关键词

/α计数/YAP∶Ce闪烁体/乏燃料后处理

Key words

plutonium/α-counting/YAP∶Ce scintillator/spent fuel reprocessing

分类

化学化工

引用本文复制引用

刘联伟,吴继宗,刘权卫,谈树苹..溶液中钚α活度的直接测量研究[J].原子能科学技术,2015,(8):1467-1472,6.

原子能科学技术

OA北大核心CSCDCSTPCD

1000-6931

访问量1
|
下载量0
段落导航相关论文