首页|期刊导航|重庆医学|全数字化乳腺摄影自动优化模式辐射剂量的比较研究

全数字化乳腺摄影自动优化模式辐射剂量的比较研究

陆云峰 张静 黄扬 陈荣生

重庆医学Issue(23):2740-2741,2744,3.
重庆医学Issue(23):2740-2741,2744,3.DOI:10.3969/j.issn.1671-8348.2013.23.015

全数字化乳腺摄影自动优化模式辐射剂量的比较研究

A study about the radiation dose of full field digital mammography in different AOP mode

陆云峰 1张静 1黄扬 1陈荣生1

作者信息

  • 1. 重庆医科大学附属第一医院放射科,重庆400016
  • 折叠

摘要

Abstract

Objective To study the method of selecting appropriate exposure parameter of AOP mode and reducing the radiation dose effective with the control of image quality .Methods The unique mammary phantom was mammography in three different AOP mode(DOSE mode ,STD mode and CNT mode) and compared the radiation dose .150 sample was selected randomly .50 sample each mode ,and recorded the radiation dose .All image reach the diagnostic standard .The mean dose of the each mode were computed and compared with others .Results whatever the phantom mammography and the sample mammography ,the mean dose in DOSE mode was lowest ,higher in STD mode ,and highest in CNT mode .The number of fiber cords ,mass and calcification fit the standard of quality control .The image of sample aslo fit the diagnostic standard .Conclusion In daily work ,we can lower the dose choosing appropriate mode according to the thickness and density of mammary .DOSE mode prior ,STD mode advocate limit to CNT mode .

关键词

乳腺摄影/辐射剂量/自动优化模式

Key words

mammography/dose/automatic optimization of parameter