检验医学与临床Issue(1):43-44,2.DOI:10.3969/j.issn.1672-9455.2013.01.020
Sysmex CA1500全自动血凝仪纤维蛋白原两种测定方法的应用比较
Evaluation of two different methods of fibrinogen assay by the automatic coagulometer of Sysmex CA 1500
摘要
Abstract
Objective To evalute two different methods of fibrinogen assay by the automatic coagulometer of Sysmex CA1500 .Methods Von Clauss method and PT-der method were used to detect 167 clinical samples .FIB was divided into 3 groups based on the results of PT-der method .The result of FIB was analyzed by statistical method . Results Compare with two methods of PT-der and Von Clauss ,there was not statistically significant difference (P>0 .05) when the concentration of fibrinogen was between 2 .0 -4 .0 g/L .When the concentration of fibrinogen was lower than 2 .0 g/L or higher than 4 .0 g/L ,the two methods had statistically significant difference (P<0 .05) .Con-clusion Von Clauss has high veracity ,it is a fast and simple method ,but the reagent is very expansive .PT-der meth-od don't any reagent ,so the cost is low .The combination of the two methods can make sure that works are good quali-ty and rapid ,improve efficiency ,and reduce the reagent cost .关键词
全自动血凝仪/纤维蛋白原/Von Clauss法/PT-der法Key words
automatic coagulometer/fibrinogen/von clauss method/PT-der method引用本文复制引用
冯彩莲,康志君,李诗韵,黄瑜璇..Sysmex CA1500全自动血凝仪纤维蛋白原两种测定方法的应用比较[J].检验医学与临床,2013,(1):43-44,2.