| 注册
首页|期刊导航|中国医疗设备|IQ·SPECT和LEHR·SPECT的心肌模型显像研究

IQ·SPECT和LEHR·SPECT的心肌模型显像研究

王猛 周翠红 金超岭 颜珏

中国医疗设备Issue(11):47-49,3.
中国医疗设备Issue(11):47-49,3.DOI:10.3969/j.issn.1674-1633.2015.11.014

IQ·SPECT和LEHR·SPECT的心肌模型显像研究

IQ·SPECT System and LEHR·SPECT System in Myocardial Perfusion Imaging:a Phantom Study

王猛 1周翠红 2金超岭 1颜珏1

作者信息

  • 1. 中日友好医院 核医学科,北京 100029
  • 2. 中日友好医院 手术麻醉科,北京 100029
  • 折叠

摘要

Abstract

Objective To investigate the inlfuence of application of IQ·SPECT (Single-Photon Emission Computed Tomography) and LEHR·SPECT on myocardial perfusion imaging.Methods The phantom of myocardial was used for comparison between IQ·SPECT and LEHR·SPECT after routine acquisition and image reconstruction. Then, the influence of application of different acquisition time on the quality of images was analyzed.Results With the extension of the acquisition time for each frame, the noise of the reconstructed image was reduced gradually, and the percentage of the radioactivity in the cooling zone was also decreased gradually. For the phantom image with same acquisition time, IQ·SPECT demonstrated superiority over LEHR·SPECT in its smaller image noise and percentage of the radioactivity in the cooling zone.Conclusion For myocardial perfusion imaging, application of IQ·SPECT could obtain higher-quality images with high-efifciency acquisition.

关键词

IQ·SPECT/LEHR·SPECT/心肌模型显像

Key words

IQ·single-photon emission computed tomography/LEHR·single-photon emission computed tomography/myocardial perfusion imaging

分类

医药卫生

引用本文复制引用

王猛,周翠红,金超岭,颜珏..IQ·SPECT和LEHR·SPECT的心肌模型显像研究[J].中国医疗设备,2015,(11):47-49,3.

中国医疗设备

OACSTPCD

1674-1633

访问量0
|
下载量0
段落导航相关论文