POCT血糖仪与全自动生化分析仪血糖测定结果的比对分析

王宾 张剑平 李耀军

医学信息Issue(37):360-360,361,2.
医学信息Issue(37):360-360,361,2.

POCT血糖仪与全自动生化分析仪血糖测定结果的比对分析

Comparative Analysis of Glucose Determination Results Using POCT Glucose Meter and Automatic Biochemical Analyzer

王宾 1张剑平 2李耀军3

作者信息

  • 1. 陕西省陇县中医医院检验科,陕西 陇县 721200
  • 2. 陕西省陇县妇幼保健院检验科,陕西 陇县 721200
  • 3. 西安医学院第二附属医院检验科,陕西 西安 710038
  • 折叠

摘要

Abstract

Objective To study test results of using the Johnson &johnson glucose meter and BS - 420 ful y automatic biochemical analyzer. Methods 81 cases patients who needed blood glucose detection in our hospital outpatient service were selected, Johnson &johnson glucose meter was used to determine peripheral blood glucose, the BS - 420 analyzer to detect serum glucose, analysis of the dif erences in test results. Results Johnson &johnson glucose meter and BS - 420 analyzer determination results were 9.45±4.28,9.26±4.12mmol/L respectively, there is no statistical dif erence (> 0.05). Conclusion Johnson &johnson glucose meter has smal volume, simple operation and fast inspection speed, can be used in the screening of outpatient emergency treatment in patients with diabetes.

关键词

血糖检测/POCT血糖仪/全自动生化分析仪/比对分析

Key words

Glucose determination/POCT glucose meter/Automatic biochemical analyzer/Comparison analysis